Flatfoot, or pes planus, represents one of the most prevalent structural deformities encountered in podiatric and orthopaedic practice. Characterised by collapse of the medial longitudinal arch, excessive hindfoot eversion, and anterior displacement of the talus relative to the calcaneus, symptomatic flatfoot can significantly impair quality of life and contribute to a cascade of proximal biomechanical dysfunction affecting the knees, hips, and lumbar spine. While conservative management remains the first-line approach, a subgroup of patients with persistent symptoms and documented structural pathology may be candidates for surgical intervention. Among the minimally invasive options that have garnered considerable attention over the past two decades is HyProCure — a subtalar extra-osseous talotarsal stabilisation device.
Pathomechanics of Flatfoot and the Role of the Subtalar Joint
To understand the therapeutic logic of HyProCure, one must first appreciate the central role of the subtalar joint in flatfoot pathomechanics. Talotarsal dislocation — partial, recurrent displacement of the talus on the tarsal mechanism — is considered by proponents of the HyProCure system to be the underlying driver of acquired flatfoot deformity. During stance and gait, the talus displaces anteriorly and medially off the sustentaculum tali of the calcaneus, collapsing the arch and producing the characteristic abducted forefoot and everted hindfoot appearance. This malalignment places pathological stress on the plantar fascia, posterior tibial tendon, spring ligament complex, and ankle mortise — structures commonly implicated in chronic pain syndromes associated with flatfoot.
Traditional surgical approaches, including calcaneal osteotomies and medial column procedures, address the downstream consequences of this malalignment but involve significant osseous disruption, prolonged recovery, and irreversible structural alteration. HyProCure offers an alternative paradigm: correcting the root cause by mechanically stabilising the sinus tarsi — the anatomical space between the talus and calcaneus — and thereby restoring normal talotarsal alignment.
The HyProCure Device: Design and Mechanism
HyProCure is a titanium stent implanted into the sinus tarsi through a minimally invasive approach. The device functions as a spacer, resisting the pathological anterior and medial displacement of the talus during weight-bearing. Unlike subtalar arthroereisis implants of earlier generations, HyProCure is positioned within the sinus tarsi rather than the subtalar joint itself, aiming to preserve physiological range of motion while limiting excessive pronatory excursion.
The implant is available in multiple sizes to accommodate variations in sinus tarsi anatomy, and sizing is typically determined intraoperatively under fluoroscopic guidance. The surgical procedure is performed through a small lateral incision, with the device inserted along the axis of the sinus tarsi. The procedure is generally completed in under 30 minutes per foot and can be performed under local or general anaesthesia in an outpatient setting. Recovery involves protected weight-bearing for a variable period, with most patients returning to normal footwear within several weeks.
Clinical Indications and Patient Selection
Appropriate patient selection is paramount to achieving satisfactory outcomes with HyProCure. The device is generally indicated for patients with flexible flatfoot deformity and documented symptomatic talotarsal dislocation who have failed a structured course of conservative management, including orthotic therapy, physical therapy, and footwear modification. It is not appropriate for rigid or fixed flatfoot deformities, where the subtalar joint cannot be manually reduced to a neutral position.
HyProCure has been studied across paediatric, adolescent, and adult populations. In children, the procedure is sometimes advocated to correct pathological pronation during skeletal development, with the premise that early structural correction may prevent secondary deformities. However, this paediatric application remains a subject of debate within the profession, given concerns about overtreatment and the natural history of juvenile flatfoot.
Contraindications include subtalar arthritis, prior subtalar fusion, tarsal coalition, and conditions associated with rigid hindfoot deformity. A thorough preoperative evaluation — including weight-bearing radiographs, gait analysis, and clinical examination — is essential to confirm flexibility of the deformity and exclude contraindications.
Evidence Base and Clinical Outcomes
The evidence surrounding HyProCure has grown substantially since its introduction in the early 2000s, though the quality of the literature remains heterogeneous. Multiple prospective and retrospective studies have reported significant improvements in pain scores, functional outcomes, and radiographic alignment parameters following implantation. Studies utilising the American College of Foot and Ankle Surgeons (ACFAS) scoring system and validated patient-reported outcome measures have consistently demonstrated meaningful reductions in pain and improvements in activity levels.
Radiographically, implantation has been associated with favourable changes in talar declination angle, calcaneal inclination angle, and lateral talo-first metatarsal angle — markers of improved arch architecture and hindfoot alignment. Gait laboratory analyses have shown reductions in peak subtalar pronation and improvements in ankle dorsiflexion kinematics postoperatively.
The reported removal rate — a meaningful surrogate for device intolerance — has varied considerably across studies, ranging from approximately 5% to over 20% in some series. Common reasons for removal include persistent sinus tarsi pain, implant migration, and failure to achieve satisfactory symptom relief. These figures underscore the importance of meticulous patient selection and surgical technique. When HyProCure is used as part of a comprehensive surgical plan that may include adjunctive soft tissue or osseous procedures, outcomes tend to be more predictable than when used in isolation for complex or severe deformities.
Controversies and Considerations
Despite growing clinical interest, HyProCure is not without controversy. Critics have questioned whether the biomechanical premise — that sinus tarsi impingement corrects talotarsal dislocation — is sufficiently supported by high-quality biomechanical research. The concept of extra-osseous talotarsal stabilisation draws on a specific theoretical model that is not universally accepted within the podiatric and orthopaedic communities.
Furthermore, the relative paucity of randomised controlled trials comparing HyProCure to alternative surgical procedures limits definitive conclusions about its superiority or equivalence. Long-term follow-up data beyond five years remain sparse, leaving questions about implant durability and late-onset complications partially unanswered.
Finally…
HyProCure represents a compelling minimally invasive option in the surgical armamentarium for symptomatic flexible flatfoot. Its ability to address the underlying talotarsal instability through a simple outpatient procedure, with rapid recovery and a meaningful evidence base supporting clinical efficacy, positions it as a viable alternative to more traditional reconstructive surgery in appropriately selected patients. Nonetheless, clinicians must approach its application with rigorous patient selection, realistic outcome expectations, and an awareness of the current limitations in the evidence base. Ongoing high-quality research — including comparative trials and long-term outcome studies — will be essential to more precisely define HyProCure’s role within the broader management algorithm for flatfoot deformity.